Display ranges from 5 x 10-9 to 2000 mbar (0.75 x 10-9 to 1500 Torr)
Readout selectable between mbar, Torr or Pascal
Adjustable switching thresholds with variable hysteresis, floating changeover contacts and visual indication of the switching status in the display
option of entering gas correction factors
Separate chart recorder outputs 0 – 10 V for each measurement channel
Compact bench top enclosures which can also be rack mounted for fitting into 19", 2 or 3 HU racks
Our combined cold cathode gauges use the inverted magnetron principle of cold cathode ionization and is combined with a pirani gauge.
By combining the technologies into a single more compact sensor, the measuring range is increased enabling a reliable pressure measurement from atmosphere to 1 x 10-9.
The on/off function of the cold cathode is controlled automatically by the integrated pirani, seamlessly eliminating any risk of running the cold cathode at too high a pressure which could lead to a reduced lifespan.
Both analogue 0-10V Linear outputs and digital communications are available.
Their high accuracy, fast response times, and robust ceramic diaphragm, make them unbeatable for measurements between 1000 and 1 x 10-4Â mbar (1 x 10-5Â Torr).